Jeio Tech ILP Personal Low-Temperature Incubators – Peltier, 14 & 48 L
Jeio Tech (Lab Companion) ILP personal incubators use Peltier (thermoelectric) cooling instead of a compressor, so they heat and cool quietly with very little vibration, noise or heat output. They control down to 5 °C (at 20 °C room temperature) and up to 40 °C, with ±0.1 °C fluctuation at 25 °C, which suits BOD-style tests, insect and plant work and sensitive samples near room temperature.
Both sizes are compact and stackable, with a stainless steel interior and shelves. Features include three stored temperatures, a wait on/off timer, adjustable air-circulation speed and defrost cycle, computer connection for data, an over-temperature limit, sensor-fault alarm and controller lock. They run on 120V and are available in Canada from Cambridge Environmental Products with CAD pricing.
Ideal For
- Near-ambient incubation (e.g. 20–30 °C)
- Vibration-sensitive samples
- Insect, plant and seed work
- Personal or benchtop use where space is tight
Why Choose the ILP Personal Incubator?
Peltier cooling
No compressor: low vibration, noise and heat output.
Heats and cools
Down to 5 °C (at 20 °C ambient) and up to 40 °C.
Stable
±0.1 °C fluctuation at 25 °C.
Compact and stackable
Stainless interior and shelves.
Available Models
Choose a Chamber Size in the selector.
| Code | Model | Volume | Interior W×D×H (in) | Exterior W×D×H (in) | Shelves (std/max) | Variation at 25 °C | Net Weight | Current (120V) |
|---|---|---|---|---|---|---|---|---|
| AAH24033U | ILP-02 | 14 L | 12.4 × 7.9 × 9.1 | 17 × 19.4 × 15.7 | 1 / 6 | ±0.5 °C | 27.5 kg (60.6 lb) | 3.4 A |
| AAH24043U | ILP-12 | 48 L | 13.1 × 13.1 × 17 | 17 × 26.2 × 24.4 | 2 / 13 | ±0.4 °C | 49 kg (108 lb) | 6.7 A |
Lowest temperature depends on room temperature: 5 °C at 20 °C ambient, 10 °C at 25 °C ambient, 15 °C at 30 °C ambient.
Which Model Should I Choose?
- A few plates or tubes, personal use: 14 L (ILP-02).
- More samples or taller vessels: 48 L (ILP-12).
- Need to go near 0 °C or above 40 °C? Choose an IL3 refrigerated incubator.
How to Use
- Place the unit with clearance for airflow, away from heat sources.
- Set temperature, fan speed and timer as needed.
- Allow time to stabilise (cooling to 5 °C takes about 60 min).
- Load samples and keep door openings short.
- Defrost and clean periodically.
Technical Specifications
| Manufacturer | Jeio Tech (Lab Companion) |
| Cooling | Peltier (thermoelectric) |
| Temperature range | 5–40 °C at 20 °C ambient; 10–40 °C at 25 °C; 15–40 °C at 30 °C |
| Fluctuation at 25 °C | ±0.1 °C |
| Variation at 25 °C | ±0.5 °C (14 L); ±0.4 °C (48 L) |
| Heating to 40 °C | 17 min (14 L); 9 min (48 L) |
| Cooling to 5 °C | 60 min |
| Controls | 3 stored temperatures, wait on/off timer, fan speed, defrost control, PC connection |
| Safety | Over-temperature limit, sensor-fault alarm, controller lock |
| Electrical | 120V, 60Hz; 3.4 A (14 L), 6.7 A (48 L) |
Specifications are from Jeio Tech catalog data and Jeio Tech distributor listings.
Selection & Handling Notes
- 120V, 60Hz, 3.4 or 6.7 A. Check the circuit rating before installing.
- Low-end temperature depends on room temperature. Keep the room cool for set points near 5 °C.
- No internal outlet or access port, per a Jeio Tech distributor.
- Electrical certification: contact us to confirm CSA/cUL status for this model before purchase if your site requires it.
Canadian Purchasing Support
Cambridge Environmental Products is a Canadian-owned laboratory supply distributor based in Ontario and has served Canadian labs since 1993. We supply Jeio Tech (Lab Companion) laboratory equipment with CAD pricing and shipping across Canada. We work with research, agricultural, environmental and teaching labs, and can help with quotes and choosing the right model. Browse our lab supplies collection for more.
Frequently Asked Questions
How cold can it go?
Down to 5 °C at 20 °C room temperature (10 °C at 25 °C, 15 °C at 30 °C).
Why Peltier?
Thermoelectric cooling has no compressor, so it runs with very little vibration and noise.
Can two be stacked?
Yes, they are designed to stack.
Is the Jeio Tech ILP Personal Incubator available in Canada?
Yes. Cambridge Environmental Products supplies it across Canada with CAD pricing.
Need Help Choosing a Model?
Tell us:
- Your set temperature
- Room temperature
- Sample volume
We'll recommend the right model.
Contact Cambridge for Product Support