CHEMetrics Glycol CHEMets Test Kit & Refill

SKU: K-4815

CHEMetrics Glycol CHEMets Test Kit & Refill

SKU: K-4815
Title
Regular price $189.30 CAD
/
  • Secure payments
Shipping calculated at checkout.

CHEMetrics Glycol CHEMets Test Kit – K-4815

The CHEMetrics K-4815 Glycol CHEMets Test Kit is a visual water test for detecting ethylene glycol and propylene glycol in cooling water, stormwater and industrial process water. It is useful for checking possible glycol contamination or leaks without purchasing a photometer.

The kit uses the Purpald-Periodate method, with measurement ranges of 1–15 ppm and 10–300 ppm as ethylene glycol using the appropriate dilution procedure. Results take approximately 17 minutes. Each kit contains supplies for 30 tests.

  • 30 tests: Complete kit or replacement ampoules available.
  • 1 ppm detection limit: Suitable for low-concentration glycol screening.
  • No instrument required: Compare the colour developed in the ampoule against the supplied comparator.
  • Extended testing ranges: Higher concentrations can be tested using specified sample dilutions.

Ideal Applications

Industrial cooling water, heat-transfer system leak investigations, airport de-icing runoff, stormwater monitoring and environmental water testing where ethylene or propylene glycol may be present.

Important: This test detects both ethylene and propylene glycol but cannot tell them apart. It is not intended to directly measure antifreeze strength, freeze protection or seawater glycol concentrations.

Choose the Complete Kit or Refill

Product SKU What's Included
Complete Glycol Test Kit K-4815 30 tests, comparator, activator solutions, 25 mL sample cup and lid, 3 mL syringe, instructions and case.
Glycol CHEMets Refill R-4815 30 replacement reagent ampoules only.


First-time buyer?
Choose K-4815. If you already own the kit, R-4815 replaces the ampoules. The refill does not replace the comparator or activator solutions, so check their condition and expiry before reordering. Distilled water is required separately.

Product Details

Technical Specifications
Manufacturer CHEMetrics
Complete Kit K-4815
Refill R-4815
Measurement Ethylene glycol / propylene glycol response
Range 1–15 ppm; 10–300 ppm with dilution
Extended Range Up to 12,000 ppm EG or 24,000 ppm PG using specified dilutions and suitable sample-measuring equipment
Minimum Detection Limit 1 ppm
Method Purpald-Periodate, visual colour comparison
Analysis Time Approximately 17 minutes
Accuracy ±1 colour standard increment under ideal laboratory conditions
Tests Per Kit / Refill 30
Storage Dark, room-temperature storage; protect from excessive heat
How Does the Glycol Test Work?

The test uses a chemical reaction that converts ethylene glycol or propylene glycol into formaldehyde. The formaldehyde reacts with Purpald reagent to produce a purple colour.

The operator compares the colour with the supplied comparator to estimate glycol concentration.

Results are reported in ppm (mg/L) as ethylene glycol. For a sample known to contain propylene glycol rather than ethylene glycol, the manufacturer instructs users to multiply the reading by two.

Follow the CHEMetrics instructions for sample preparation, reagent additions, timing and any required dilution.

Testing Limitations and Interferences

The Purpald-Periodate method is useful for glycol screening, but it has limitations:

  • It cannot distinguish ethylene glycol from propylene glycol.
  • Formaldehyde and certain other aldehydes can cause falsely elevated results.
  • Strong oxidizers and reducing agents can interfere with the test.
  • Methanol at percentage concentrations can produce falsely elevated readings.
  • High iron concentrations, coloured samples or turbidity can make colour comparison difficult.
  • Samples should be below 40°C for best results.
  • Seawater is not a suitable sample matrix.
  • Diethylene and triethylene glycol are not measured by this method.

If your facility needs to identify the specific glycol present or quantify glycol for regulatory reporting, confirm whether an appropriate laboratory analytical method is required.

Shelf Life, Storage and Reordering

CHEMetrics currently lists the R-4815 refill shelf life as six months from the date of manufacture. Confirm the expiry or lot dating of the product supplied, as manufacturer documentation has also specified five months.

Store reagents in the dark at room temperature and avoid excessive heat. Visible crystals in an unused reagent ampoule may indicate deterioration.

Before purchasing several refills, consider your testing frequency and how many tests you can complete before the reagent expires.

For ongoing programs, verify that the comparator and activator solutions remain usable. Replacement reagents and accessories may need to be ordered separately.


Frequently Asked Questions

Does this kit test ethylene glycol and propylene glycol?

Yes. Both respond to the test chemistry, but the kit cannot identify which glycol is present. Results are read as ethylene glycol equivalents.

Can I use it to detect glycol leaks in cooling water?

Yes. The kit is suitable for screening compatible cooling-water samples for glycol. Check expected concentration and possible chemical interferences before testing.

Can I test antifreeze concentration directly?

This is a low-concentration water test. Although dilution extends the measurement range, it is not intended to directly determine antifreeze percentage or freezing protection in concentrated coolant. A suitable refractometer or other coolant-specific method may be more appropriate.

How many tests are included?

The K-4815 complete kit and R-4815 refill each provide 30 tests.

Does the refill include the comparator and activator solutions?

No. R-4815 contains 30 replacement ampoules. The comparator and activator solutions are supplied with the complete K-4815 kit or sourced separately when required.

Does this test require a photometer?

No. Results are determined visually by matching the developed colour against the supplied comparator.

Can the kit be used for airport runoff or stormwater?

Yes. CHEMetrics identifies aircraft de-icing operations and stormwater among the intended applications. Sample conditions and potential interferences should be evaluated before use.

Is this suitable for regulatory compliance testing?

Do not assume this visual kit meets a particular regulatory reporting requirement. Confirm the required analytical method, reporting limit and sampling procedures with the relevant authority or laboratory before using results for compliance purposes.


Manufacturer Documents and Related Products

CHEMets Self-Filling Ampoules

This general demonstration shows how CHEMets self-filling ampoules work. Follow the specific K-4815 instructions for glycol testing.

Need Help Choosing a Glycol Test Kit?

Not sure whether the CHEMetrics K-4815 is right for your water samples? Cambridge Environmental can help Canadian industrial water, wastewater and environmental testing customers review the measurement range, sample requirements and refill options.

Tell us what type of water you are testing, the expected glycol concentration and whether you need a complete kit or refill. We can help you select the appropriate product.

Contact Cambridge for Glycol Testing Support