Keep your sensitive vaccines and pharmaceuticals safe with Fridge-Tag® 2L Temperature Data Loggers. No need for software as it automatically generates a daily summary of temperature and 8 weekly graphs. With temperature readings recorded every 5 minutes, you can trust its compliance with FDA regulations 21 CFR Part 11.
NIST traceable with a 3 year certificate, this logger offers bio-safe glycol/water probe medium for accurate readings. Monitor with ease using the current temp display and 30-day status screen. Meets FDA requirements and records 43,200 observations monthly for reliable data. Get monthly reports and comply with regulations effortlessly. The Bio-Safe glycol filled bottle and external thermometer probe buffer temperature changes. Meets CDC guidelines for storing and monitoring vaccines.
Included: Data logger, 3yr NIST certificate, Vial & Probe
SPECIFICATIONS:
External dimensions (L x W x H) |
128 x 75 x 14.5mm |
Size of the LCD display |
95 x 27mm |
Weight |
96g |
Storage condition (inactive) |
-30°C to +60°C / -22°F to +140 °F |
Operating temperature |
-30°C to +55°C / -22°F to +131 °F |
Display visible |
-10°C to +55°C / +14°F to +131 °F |
Alarm temperature range |
-30°C to +50°C, in 0.1°C increments / -22°F to +122°F |
Alarm time |
1 Minute to 23h 59min |
Temperature measurement interval |
Every minute |
Accuracy of temperature measurement |
± 0.8°C (-25°C to -10°C) ± 1.44°F (-13°F to +14°F) ± 0.5°C (-10°C to +40°C) ± 0.9°F (+14°F to +104°F) ± 0.8°C (+40°C to +55°C) ±1.44°F (+104°F to +131°F) |
Accuracy of time |
±30 minutes / year |
Calibration |
Every device is calibrated / NIST/ILAC traceable |
Date/Time programming |
Date and time will be set when device is activated (see operation manual) |
Protection class |
IP67 |
Operating lifetime |
up to 3 years – expiry date printed on label |
Manipulation |
Fridge-tag® 2 cannot be manipulated or reset without destroying it. |